The CODMAN CERTAS Tool Kit is provided as a
non-sterile product.
CAUTION: Do not sterilize the CODMAN CERTAS
Tool Kit.
Disinfecting the CODMAN CERTAS
Tool Kit
After use, wipe the areas of the tools that contact
the patient with a combination wipe (Quaternary
Ammonium/Isopropyl Alcohol). Follow Steps 1
through 4 below.
1. Thoroughly wet the surfaces with a wipe.
2. Keep the surfaces wet for 2 minutes. Use as
many wipes as needed to keep the surfaces
wet for the entire 2 minutes.
3. Allow the wetted surfaces to air dry.
4. Inspect the components to ensure that all soil,
blood, or debris has been removed. If needed,
repeat Steps 1 through 3 and inspect again.
Visually inspect all parts after disinfection to
ensure markings remain. Do not use the Tool
Kit components if markings are compromised.
Contact your local Integra sales representative
for a replacement kit.

Instructions for Use

Note: Additional training materials are available
from your local Integra sales representative.
SECTION A: Pre-implantation
Adjustment Procedure
Adjust the performance setting before implanting
the CERTAS
is performed before the sterile inner package of
the valve is opened.
1. Place the valve sterile package on a non-
metallic surface so the clear portion of the
package faces up. Place the Adjustable
Height Locator Tool in the circular grooves
Plus Programmable Valve. This step
5