of the package so that the arrow on the tool
points in the same direction as the arrows
molded into the plastic package (See Figure 2).
Ensure that the Adjustable Height Locator
Tool is fully seated in the grooves. If necessary,
maintain hold of the Adjustable Height
Locator Tool throughout the procedure to
keep it firmly in the grooves of the package.
CAUTION: Failure to maintain a fully-seated
Locator Tool above the valve mechanism
could result in an inaccurate indication of
the performance setting.
Note: Do not use the Low Profile Locator
Tool for programming the valve in the package,
as it does not contain a base that will fit into
the circular grooves of the packaging.
2. Fully seat the Indicator Tool into the Adjustable
Height Locator Tool so that the red markings
are aligned (Figure 3). You may hear a click when
the tool seats; the Indicator Tool will not rotate
within the Locator Tool when properly seated.
CAUTION: When reading the performance
setting with the Indicator Tool, make sure
that the Adjustment Tool and any other
magnetic devices are at least 36 cm (14 in.)
away from the Indicator Tool. Failure to do
so can result in an inaccurate indication of
the performance setting.
3. When the number dial on the Indicator Tool
stops moving, observe the number and its
position within the window to determine the
performance setting. A single number within
the window, with full purple background
and both gray side lines visible (Figure 12),
indicates the performance setting of the valve
when the tools are properly aligned. Once
the valve performance setting has been read,
remove the Indicator Tool.
6