10. EMC TABLE
The Compex stimulator needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided.
• RF emissions CISPR 11 Group 1
• RF emissions CISPR 11 CLASS B
• IMMUNITY test level
GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC EMISSIONS
The Compex stimulator is intended for use in the electromagnetic environment specified below. The customer or
the user of the Compex stimulator should assure that it is used in such an environment.
EMISSION TESTS
Radiated Emissions CISPR 11
Conducted Emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations IEC 61000-3-3
COMPLIANCE
Group 1
Class B
Class A
Complies
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
The Compex uses RF energy only
for its internal function. Therefore,
its RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
The Compex is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-volt-
age power supply network that
supplies buildings used for domestic
purposes.