GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC IMMUNITY
The Compex stimulator is intended for use in the electromagnetic environment specified below. The customer or
the user of the Compex stimulator should assure that it is used in such an environment.
IMMUNITY TEST
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast transient/
burst IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply
input lines
IEC 61000-4-11 a
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
26
IEC 60601 TEST LEVEL
±8kV contact
±2, 4, 8, 15kV air
±2kV for power supply
lines
±1kV for input/output lines
±0.5, 1kV Line to Line (or
Neutral)
±2kV line(s) to earth
0% UT
(100% dip in UT) for 0,5
cycle
0% UT
(100% dip in UT) for 1 cycle
70% UT
(30% dip in UT) for 25
cycles
0% UT
(100% dip in UT) for 5 sec
30 A/m
a UT is the a.c mains voltage prior to application of the test level.
COMPLIANCE LEVEL
±8kV contact
±2, 4, 8, 15kV air
±2kV (power lines)
Not Applicable (I/O lines)
±0.5, 1kV Line to Line
Not Applicable (Line to
Earth)
0% UT
(100% dip in UT) for 0,5
cycle
0% UT
(100% dip in UT) for 1 cycle
70% UT
(30% dip in UT) for 25
cycles
0% UT
(100% dip in UT) for 5 sec
30 A/m
ELECTROMAGNETIC
ENVIRONMENT -
GUIDANCE
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Mains power quality
should be that of a typical
domestic environment.
Mains power quality
should be that of a typical
domestic environment.
Mains power quality
should be that of
a typical domestic
environment. If the user
of the Compex stimulator
requires continued
operation during power
mains interruptions, it
is recommended that
the Compex stimulator
be powered from an
uninterrupted power
supply or a battery.
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
domestic environment.