GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC IMMUNITY
The Compex stimulator is intended for use in the electromagnetic environment specified below. The customer or
the user of the Compex stimulator should assure that it is used in such an environment.
IEC 60601 TEST
IMMUNITY TEST
LEVEL
3 Vrms b
150 kHz to 80
6 Vrms a
in ISM bands
Conducted RF IEC
710 to 780 MHz
61000-4-6
5.24 to 5.785 GHz
Radiated RF IEC
10 V/m
61000-4-3
80 MHz to 2.7
27 V/m
385 MHz
28 V/m
450 to 930 MHz
1.72 to 2.45 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the Compex stimulator is used
exceeds the applicable RF compliance level above, the Compex stimulator should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
COMPLIANCE
LEVEL
3 Vrms
MHz
6 Vrms
9 V/m
9 V/m
10 V/m
GHz
27 V/m
28 V/m
and reflection from structures, objects and people.
relocating the Compex stimulator.
ELECTROMAGNETIC ENVIRONMENT -
GUIDANCE
Portable and mobile RF communications
equipment should be used no closer to any part of
the Compex stimulator, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2,5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance
in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
EN
27