ORBIMAT 300 SW
No.
Error message
39
External storage medium not
found!
Possibly no USB stick is
inserted in the USB port
40
Error when saving log
41
Procedure file cannot be
opened!
42
Log file cannot be opened!
43
Some currents are set too
high.
They cannot be achieved with
a 115 V mains voltage!
44
An error occurred while read-
ing the file FAILURES.TXT!
45
No weld head connected!
46
Incorrect weld head con-
nected!
47
The procedure contains pa-
rameters that lie outside the
limit values for the selected
weld head!
48
Motor current
>> Collision recognition
Norm. motor current
Motor current time
49
Electrode replacement warn-
ing
50
Calibration error/fault
("Calibration aborted due to
execution error"; "Calibration
aborted due to operational er-
ror"; "Calibration aborted due
to parameter")
51
Error while copying proce-
dures or logs
52
Printing error. Error during
communication with the
printer
53
Settings incorrect
(14.01.21) OW_ORBIMAT_300SW_BA _852060202_00_EN
Explanation/Elimination
Access to external USB stick not possible.
X Check whether the USB stick is recognized by the system.
X Remove the USB stick, wait briefly and insert it again. Possibly use a differ-
ent port.
X Check whether the USB stick is defective or has an unreadable format.
X Contact Orbitalum service if the problem cannot be resolved.
Software conflicts due to damaged files or files with unreadable contents.
The message is followed by the error in "plain text" (see Messages 18 through
29) or as a two-digit number.
Machine connected to mains network of less than 200 V AC. The maximum weld
current is limited to 120 A. Procedure with higher weld current has been started.
Internal error log ("Failures.txt" file) is not present or is damaged.
Procedure started, no weld head connected to the machine.
Weld head connected?
•
Weld head not encoded
•
Voltage supply to weld head defective. The operating buttons in the weld
head are not working.
X Contact the Orbitalum service.
The currently connected weld head is not the same as the weld head in the
started procedure.
X Select the connected weld head in the procedure.
– or –
X Connect the weld head selected in the procedure to the machine.
Are the weld head models identical, but the message appears nevertheless?
X Contact the Orbitalum service.
Weld head changed in the procedure. Weld head does not match the param-
eters.
The weld head list of the machine contains the parameters of the weld heads
which can be connected to the machine, e.g. the maximum rotation of the rotor,
the maximum pipe diameter to be welded and the maximum permissible cur-
rent.
X Adjust the parameters of the procedure to the selected weld head.
Transmission sluggish.
X Check the head.
X If necessary, repair the transmission.
X Replace the electrode.
X Execute calibration again or restart the machine.
X Check whether write protection is active. The drive is possible not ready.
X Check the connection to the printer.
X Check the network connection.
X Refresh the printer list.
X Check the entries.