5.5 Changing the STATIM 2000 Air Filter
The filter should be replaced every six months in order to maintain an
adequate supply of clean air during the air drying cycle.
To change the filter, follow these steps (see Figure 11):
1.
Turn the power switch at the back of the unit OFF.
2.
Remove and retain the screw
to the back of the compressor.
3.
Remove and discard the old foam air filter (1).
4.
Install the new filter
5.
Secure the filter plate
using the screw
5.6 Changing the Bacteria Retentive Air Filters
The filters should be replaced every six months or after 500 cycles to
maintain an adequate supply of clean air during the air drying cycle.
To change the bacteria retentive air filter on the STATIM 2000 and 5000,
follow these steps (see Figures 11 & 12):
1.
Power the STATIM OFF.
2.
Disconnect tube A
the filter from the filter bracket (6).
As you remove the filter from the bracket, note the
orientation of the arrow mark on the filter.
3.
When the filter is free of the bracket,
carefully disconnect tube B
4.
Before installing the replacement bacteria retentive filter
(SciCan order no. 01-102119S) check that the arrow
mark on the filter matches the direction of the arrow on the bracket.
Push the left hand filter fitting into tube B (7).
5.
Gently press the replacement filter into the filter bracket (6).
The arrow mark of the filter should be facing out and pointing to the left.
6.
Re-connect tube A
SciCan Ltd.
Maintenance
securing the filter plate
(3)
(SciCan part no. 01-100207S).
(1)
to the back of the compressor
(2)
retained during the disassembly procedure.
(3)
from the bacteria retentive filter
(4)
from the filter.
(7)
to the right hand filter fitting.
(4)
(2)
and remove
(5)
(5)
18
STATIM 2000/5000 Cassette Autoclave Operator Manual
STATIM 2000
5
6
4
7
1
2
3
STATIM 5000
5
4
7
6
Figure 11
Figure 12