5.7 Replacing the Cassette Seal

To ensure optimum performance of your STATIM cassette autoclave, change
the cassette seal every 500 cycles or every six months, whichever comes first.
Replacement seals are available from SciCan (order number 01-100028S
for STATIM 2000 and 01-101649S for STATIM 5000).
To change the cassette seal, follow these steps:
1.
Place the cassette lid and the new seal on a clean work surface.
Examine the position of the old seal in the cassette lid and arrange the
new seal in the same orientation, next to the lid.
2.
Remove the old seal and discard. Clean any residue out of the seal
channel and flush out the channel with distilled water
3.
Lubricate the new seal with the liquid seal lubricant provided
4.
Insert the rounded edge of the seal under the round lip of the lid. Align
the holes in the new seal with the holes in the lid
At every corner and at the holes in the lid, two square nibs should
NOTE:
be visible. The nibs should fit flush with the lid's outer surface.
5.
Ensure the seal is completely inserted. Feel around the periphery to
ensure the seal is securely in place.
During a cycle, steam may appear between the lid and the tray.
NOTE:
If this persists, remove the cassette and check that the seal is
correctly installed
Be careful. The metal parts will be hot, and the cassette may contain hot steam.

5.8 Maintaining Fluid Levels

Use only steam-process distilled water containing less than 5 ppm
total dissolved solids (having conductivity of less than 10 µS/cm) in the
STATIM. To fill the reservoir, remove the cap from the top of the unit and
fill the reservoir. We recommend using a funnel to minimize spills. Each
time you refill the reservoir, empty the waste bottle and refill with water
to the MIN line. Empty the waste bottle often to avoid unpleasant odors
and discoloration of the contents. A low-level chlorine-free disinfectant,
prepared according to the manufacturer's instructions, may be added to
the waste bottle.

5.9 Reading Water Quality

1.
Power up unit while pressing the STOP button to access the User
Setup menu.
2.
Using the UNWRAPPED and WRAPPED buttons,
scroll to Water Quality and select it by pressing the
RUBBER AND PLASTIC button.
SciCan Ltd.
Maintenance
.
(13 d)
CD = Conductivity
X X = micro S. value
(13
a).
(13
b).
(13
c).
yyy = engineering value
z. z = parts per million value
19
STATIM 2000/5000 Cassette Autoclave Operator Manual
Figure 13 a
Figure 13 b
Figure 13 c
Figure 13 d