6.2 Installing the SciCan Data Logger
SciCan's USB Data Logger can record and store cycle information onto
a mass storage device (MSD) such as a USB Flash Drive or SD memory
card.
These instructions are for units with a 9-pin communication (RS232) port
located at the back of the unit. Follow the steps below before connecting
the Data Logger.
User Setup Menu
Please ensure Time/Date and Unit ID are set as per Section 3.6 and
Section 3.8. To install the SciCan Data logger, access the user setup
menu by powering up the unit while pressing the
Follow the instructions below to complete installation steps.
STEP 1: Selecting the USB Flash/MSD Option
>RS232
1.
Power up unit while pressing the STOP button to access the User
Setup menu.
2.
Using the UNWRAPPED and WRAPPED buttons, scroll to RS232 and
select it by pressing the RUBBER AND PLASTIC button.
3.
From the RS232 menu, use the UNWRAPPED and
WRAPPED buttons to scroll down to the USB/Flash MSD option
and press the RUBBER AND PLASTIC button to
select and return to the User Setup menu.
STEP 2: Setting the Serial Port Bitrate to 9600
1.
From the User Setup menu, use the UNWRAPPED and WRAPPED
buttons to scroll to the Serial Port Bitrate menu and select it using the
RUBBER AND PLASTIC button.
2.
From the Serial Port Bitrate menu, use the UNWRAPPED and
WRAPPED buttons to scroll down to 9600 and press the
RUBBER AND PLASTIC button to select and
return to the User Setup menu.
STEP 3: Setting the Printer user Character (i.e. 134°C)
1.
From the User Setup menu, use the UNWRAPPED and WRAPPED
buttons to scroll to the Printer user ° char menu and select it using the
RUBBER AND PLASTIC button.
2.
From the Printer user ° char menu, using the UNWRAPPED button to
increase the displayed value by one, and the WRAPPED button to
increase that value by ten, enter the value 32 [0x20]
and press the RUBBER AND PLASTIC button to
accept and return to the User Setup menu.
SciCan Ltd.
The Communication Port
button.
STOP
>RS232
9600
32 [0x20]
22
RS232
STATIM 2000/5000 Cassette Autoclave Operator Manual