6.3 Cycle Printout Overview
Model: STATIM 2000
Unit Identifi er: autoclave has
been set up as number 000
Time/Date:
10:47 am 25th October 2013
Cycle Name
and Parameters:
UNWRAPPED
for 135°C/3.5 min
Warm up complete:
start of the conditioning
phase is 1:19
(see cycle graph–
'A' phase complete,
start of 'B' phase)
Temp./Press. & Time
printed at 30 second
intervals during
sterilization ('D' phase).
SciCan Ltd.
The Communication Port
STATIM 2000
UNIT#: 000
CYCLE NUMBER
10:47
UNWRAPPED
CYCLE START
CONDITIONING
PRESSURIZING
STERILIZING
VENTING
AIR DRYING
CYCLE COMPLETE
Cycle completion time: 16:42.
24
S204R616
000002
10/25/2013
3. 5 MIN.
0:00
1: 19
1:27
2:27
2:27
2:57
3:27
3:57
4:27
4:57
5:27
5:57
5:57
6:42
16:42
STATIM 2000/5000 Cassette Autoclave Operator Manual
Software: S204R616
Cycle counter: the number
of cycles having been run
on the unit = 2
Cycle clock: starting at 0:00
Start Time of Pressurization:
1:27 (start of 'C' phase)
Start Time of Sterilization:
2:27 (start of 'D' phase)
Time Venting started:
5:57 (start of 'E' phase)
Time Air Drying started:
6:42 (start of 'F' phase)