Descargar Imprimir esta página

Alarm Signal Generation - Gima VITAL PRO-SNET Manual Del Usuario

Ocultar thumbs Ver también para VITAL PRO-SNET:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Medium Priority:
HR over range
High Priority:
Low battery
SpO
over limit
2
SYS over limit
DIA over limit
MAP over limit
Temp over limit
PR over limit
HR over limit
RR over limit
Unable to detect SpO
2
Unable to detect HR
Timing:

5.2 Alarm Signal Generation

When there is alarm condition, the monitor generates alarm signal with visual indications (which are shown by two ways:
LED indicator with different color and textual message display) and audible indication.
Visual Alarm Indication
The flashing rates for the three categories of alarms are shown in the table below.
LED Indicator Color
Red flashing
Yellow flashing
Yellow light
Refer to Chapter 11.2 Alarm Information for detailed alarm message descriptions.
Audible Alarm Indication
The audible alarm has different tone pitch and on-off beep patterns for each priority category. These are summarized in
the table below.
Alarm Category
High priority alarm
Medium priority alarm
Low priority alarm
Note: Visual alarm indication can not be suspended or removed. Audible alarms may be decreased in volume or silenced
as described.
Alarm Category
High priority alarm
Medium priority alarm
Low priority alarm
Table 5.1
Tone Pitch
Beep Chain
~400Hz
10 beeps pause 3 sec.
~500Hz
3 beeps pause 5 sec.
~500Hz
Single beep
Table 5.2
Flashing Rate
2 Hz
0.5 Hz
Constant(on)(non-flashing)
42

Publicidad

loading