11.1 Prompt information explanations

Mute C-D: XXX seconds
NIBP C-D: XXX seconds
TOUR C-D: XXX seconds
Probe off
PR over limit
SpO
over limit
2
SYS over limit
DIA over limit
MAP over limit
NIBP error 1#
NIBP error 2#
NIBP error 3#
NIBP error 4#
Cuff error
NIBP error 5#
Air leak
NIBP over range
Over motion
Over pressure
NIBP timeout

Chapter 11 Appendix

Alarm silence count down: XXX seconds
NIBP auto measuring cycle count down: XXX seconds
Hemostat alert count down: XXX seconds
SpO
probe is off form the patient or disconnected from the device
2
PR value exceeds the high/low alarm limit
SpO
value exceeds the high/low alarm limit
2
Systolic pressure value exceeds the high/low alarm limit
Diastolic pressure value exceeds the high/low alarm limit
MAP value exceeds the high/low alarm limit
Sensor or other hardware error
Very weak signal because of the cuff, or the patient has very weak pulse
Blood pressure amplifier overflow due to excessive movement
Leaking during the pneumatic device testing
Cuff is not wrapped correctly, or is not connected
Hardware fault of NIBP module
Pneumatic part, tube or the cuff leak air
The measurement range exceeds 255mmHg (for neonates: over 135 mmHg)
Excessive movement or noise during inflation and measurement. Another
measurement will be taken.
Cuff pressure exceeds the safety limit value of software. (limit for adults:
290mmHg; limit for pediatric: 145mmHg)
Or caused by cuff extrusion or flapping cuff with force.
Adult measurement is more than 120 seconds, neonate measurement is more
than 90 seconds.