11.2 Factory Default Alarming Values and Setup Range
The factory default alarming value:
Parameter
HR
SYS
DIA
MAP
SpO
2
Pulse rate
TEMP
The high and low limits setting range:
Mode
Parameter
High limit
HR
Low limit
High limit
SYS
Low limit
High limit
DIA
Low limit
High limit
MAP
Low limit
High limit
SpO
2
Low limit
High limit
Pulse rate
Low limit
High limit
TEMP
Low limit
Note: it's limited for the high/low alarm setting value, which ensure that the high-limit setting value will not less than (or
equal to) that of the low-limit setting value. See the above table for detailed setting step.
Mode
Adult
High limit
180bpm
Low limit
40bpm
High limit
180mmHg
Low limit
60mmHg
High limit
120mmHg
Low limit
50mmHg
High limit
160mmHg
Low limit
50mmHg
High limit
100%
Low limit
90%
High limit
180bpm
Low limit
40bpm
High limit
39.0℃
Low limit
35.0℃
Adult
(1~350) bpm
(0~349) bpm
(30~280) mmHg
(29~279) mmHg
(11~232) mmHg
(10~231) mmHg
(21~242) mmHg
(20~241) mmHg
1~100%
0~99%
(1~300) bpm
(0~299) bpm
(0.1~60) ℃
(0~59.9)℃
Pediatric
200bpm
50bpm
130mmHg
50mmHg
90mmHg
40mmHg
110mmHg
40mmHg
100%
85%
200bpm
50bpm
39.0℃
35.0℃
Pediatric
(1~350) bpm
(1~350) bpm
(0~349) bpm
(0~349) bpm
(30~200) mmHg
(30~135) mmHg
(29~199) mmHg
(29~134) mmHg
(11~150) mmHg
(11~100) mmHg
(10~149) mmHg
(10~99)mmHg
(21~165) mmHg
(21~110) mmHg
(20~164) mmHg
(20~109)mmHg
1~100%
1~100%
0~99%
0~99%
(1~350) bpm
(1~350) bpm
(0~349) bpm
(0~349) bpm
(0.1~60) ℃
(0.1~60) ℃
(0~59.9)℃
(0~59.9)℃
65
Neonate
220bpm
50bpm
110mmHg
50mmHg
90mmHg
30mmHg
100mmHg
30mmHg
100%
85%
220bpm
50bpm
39.0℃
35.0℃
Neonate
Setting Step
1bpm
1bpm
1mmHg
1mmHg
1mmHg
1mmHg
1mmHg
1mmHg
1%
1%
1bpm
1bpm
0.1℃
0.1℃