4.10 Alarm Settings
Press the alarm silence "
There are 3 options in total:
² Alarm sound is enabled, which is the default status..
² Short term alarm silence (120 seconds): a short press of the alarm silence key will display the red icon "
lower screen and the message "silence count-down time 120". At this time, the alarm silence indicator on the left
side of the alarm silence key will be lit. The device will mute the alarm sound temporarily for 2 minutes, but keep
the visual alarm flashing. When the time (120s) is up, the alarm silence will be de-activated automatically, the red
icon "
" will disappear as well, and the alarm silence indicator will be dark.
² Long term alarm silence: long press the alarm silence key will be display the red icon "
and the alarm silence indicator on the left side of the alarm silence key will be lit. The device will mute the alarm
sound on an ongoing basis but keep the visual alarm flashing untill a new type of alarm event is detected. The alarm
silence status will then be terminated automatically and the alarm sound will resume, the red icon "
disappear as well, and the alarm silence indicator will be dark.
Note: If the current status is alarm silence, a long or short press of the alarm silence key can de-activate the alarm silence
function.
4.11 Data Uploading
When connected to a computer via USB, the device enters into data uploading mode, as shown in figure 2.25.
In data uploading mode, the device will automatically stop SpO
verification and air leakage checks etc., and all key operations will be disabled except power "
" key to set the alarm sound status.
Figure 4.25 Data uploading screen
measurement, NIBP measurement, Hemostat, pressure
2
40
" on the
" on the lower screen
" will
" key.