Recommended separation distances between portable and mobile RF
communications equipment and The equipment or system-
for EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING
Gima Vital Pro Monitor is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the equipment or system can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the equipment or system as recommended below, according to
the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be determined using the equation applicable to the frequency of the transmitter, where p is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2:These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Table 4
Separation distance according to frequency of transmitter
m
150kHz to 80MHz
0.12
0.38
1.2
3.8
12
80MHz to 800MHz
0.12
0.38
1.2
3.8
12
53
80MHz to 2,5GHz
0.23
0.73
2.3
7.3
23