13. Error Messages and Fault Clearance
Messages are displayed in plaintext and are self-explanatory.
Error Message / Error
"Dry Strip"
"Wrong Strip"
"Wrong Position"
"No Paper"
"Battery Low"
"Instrument doesn't
start"
In case a fault cannot be cleared by the aid of the instructions above, please contact your local
distributor or the MACHEREY-NAGEL Service.
Manual URYXXON
100
Cause
The test strip wasn't dipped
completely
A wrong test strip has been
detected (wrong type)
The strip hasn't been pushed into
the strip retainer far enough
Paper roll empty or printer flap
open
Batteries are low
Power supply not installed or
defect
Relax EN, V 6.01 / 03.22
®
Solution
Repeat measurement with a
new strip
Use correct test strips
New measurement, place strip
in right position
Replace paper and close
printer flap
Exchange batteries or use
power pack
Check whether all connections
are plugged in and whether
the power socket is
functioning