11. Service menu
The URYXXON
Relax has a password protected service menu. To enter the service menu
®
press the touch-screen three times during the self test after turning the equipment on (Display
25). Upon request input the PIN "1234".
A selective list with different menu items appears.
11.1 How to reset the system (Load Default)
Select "Load default" from the service menu. The instrument will be reset to delivery status.
All settings modified by the user will be cleared!
11.2 How to control the LOT-control (LOT activate)
Select "LOT activate" from the service menu to activate the LOT-control. This will minimize
the risk of using expired test strips. You will be asked for the LOT of strips currently used.
The system will give a warning when the test strips are expired. After 100 measurements the
system will ask for the LOT-number of the next tube.