9. Equipment Settings
Enter the main menu by pressing the menu key
(Display 22 + Display 23).
MENU
Memory
Check mode
Settings
Display 22: Main menu
9.1 How to modify strip settings
In the "Strip" menu, settings for result displaying and the test strip type can be chosen.
9.1.1 Type
The URYXXON
Relax is in principle able to analyze also non-medical strip types. When
®
the equipment has data for various approved strip types, the strip type can be chosen in a
selective list.
9.1.2 Units
The URYXXON
Relax can report the results in different units:
®
• Conventional (e.g. 10 mg/dL)
• SI (e.g. 56 mmol/L)
• ARB, Plus-System (e.g. +++)
• Conventional + ARB
• SI + ARB
Choose the desired unit from the selective list. For the strip parameter Microalbumin the results
can only be shown in conventional units.
9.1.3 Order of Parameters
The output order of the parameters may be customized via a selective list. The parameters
need to be selected in the desired order and confirmed by pressing
parameter the equipment will ask whether the setting is to be saved. Save by pressing
or go back to the preprogrammed order by pressing
for the strip parameter Microalbumin is not possible.
9.1.4 Test Strip LOT
The LOT-administration of URYXXON
11.2 "How to control the LOT-control (LOT activate)".
Manual URYXXON
90
Relax is deactivated in the default setting. See Chapter
®
Relax EN, V 6.01 / 03.22
®
to reach the "SETTINGS" display press
SETTINGS
Strip
Password
Printer
Sound
Autostart
Battery
Language
Date/Time
Interface
Display 23: Settings
. A change of the parameter order
. After the last