1000 Hz with Reflex – Basic (Stimuli @ 1 kHz) license
o
500 Hz, 1000 Hz, 2000 Hz and 4000 Hz with Reflex – Plus
o
(Stimuli @ 0.5, 1, 2, 4 kHz) license.
-
Pure-tone audiometry test (with Tonal Audiometry license).
Other accessories available are a dedicated docking station and a portable
thermal printer. Refer to chapter 2.2 for more details.
1.6
U
SE CASES
The Timpani can be used to conduct automatic tympanometry tests at low
frequency (226 Hz) and at high frequency (1000 Hz, only with 1 kHz Probe Tone
license) and ipsilateral acoustic reflex tests (only with Reflex – Basic or Reflex –
Plus license). Moreover, it is also possible to perform the audiometry test using
pure tones by activating the Tonal Audiometry license.
These tests must be conducted in a particularly quiet environment in order to
avoid artifacts.
The Timpani tympanometer is intended for use by persons who have a detailed
knowledge of the procedures associated with the tests supported by the
instrument; the operator must therefore be either an audiometrist (or a technician
having the requisite levels of audiological knowledge) or a medical practitioner
in possession of specific skills (ENT or audiology specialist).
1.7
W
ARNINGS AND
To ensure correct and safe use of the device, the following precautions must be
observed.
1.7.1
General precautions
Make certain that the required ambient conditions are met (during
transport, storage and operation) as indicated in Appendix A.
The device will not be protected if exposed during use to flammable
anesthetic gases or similar products. Risk of explosion.
Avoid installing and using the device near sources of strong
electromagnetic fields: these could interfere with the operation of the
equipment.
Use only original accessories supplied by Inventis, unless specifically
indicated otherwise.
Use only medical grade power adapters, certified to IEC 60601-1
standard. For further information see Appendix A.