Problem
The instrument
does not switch on
The display
remains blank
(LED on)
Battery does not
recharge
A test cannot be
accessed
message:
"Hardware error"
message:
"Serious error"
46
Possible cause
Battery flat
Instrument in stand-by
Display damaged
USB cable damaged
Adapter damaged
Instrument not positioned
correctly in the docking
station
Battery damaged
Optional test not activated
Non-fatal internal error
Fatal internal error
Solution
Connect the instrument to a power
source and switch on the device
Touch the screen or press the power
button
Contact Inventis service department
or dealer
Change the USB cable (USB A –
mini B standard)
Contact Inventis service department
or dealer
Check the positioning of the
instrument
Check that the contacts are clean
Check connections
Replace the battery - Contact
Inventis service department or
dealer
Contact the Inventis technical
service department to obtain the
license, supplying the serial number
of the device
Press OK to continue; if the
problem persists, contact the
Inventis service department
Restart the instrument; if the
problem persists, contact the
Inventis service department