Manufacturer's indications and declaration – electromagnetic immunity
Function requiring verification to
exclude unacceptable risks
Sound generator operating correctly
Tympanometry on test cavity, conducted
correctly in normal operating conditions
Contains transmitter module in compliance with ETSI EN 301 489-1 and ETSI
EN 300 328 standards
The device emits radio frequencies in the 2.4 GHz band, class 1
Contains transmitter module in compliance with ETSI EN 301 489-1 and ETSI
EN 300 328 standards
50
Pass/fail acceptance criteria
No unwanted sound from transducers exceeding 80 dB; lockup or
restart of the device is acceptable
Flat tympanometry curve, ECV indication equivalent to nominal
value of cavity +/- 0.1ml
ESD: presence of artifacts in tympanometry recognizable by
skilled professional, HW error, lockup or restart of device — all
these are acceptable