Having cleaned the channels thoroughly, the tip of the probe must be refitted.
Offer the tip of the probe to the main body, being careful to align the guide pin
A presented by the body with the hole B located in the tip, as illustrated in the
figure below. Retighten the screw collar.
Clean the outer surface of the probe using a lint-free soft cloth moistened with
water and mild detergent; if the probe is to be sanitized, moisten the cloth with a
3% solution of hydrogen peroxide.
Do not dip the probe or any part of the probe in liquids of whatever
nature
In the event of the probe being broken or malfunctioning, contact an Inventis
service technician. The replacement of the probe must be entrusted exclusively
to Inventis or to a service technician authorized by Inventis. If the probe is
replaced, it must be calibrated before being used with the instrument.
10.4 C
LEANING THE INSTRUMENT
Clean the instrument using a lint-free soft cloth moistened with water and mild
detergent; if it is to be sanitized, moisten the cloth with a 3% solution of hydrogen
peroxide.
38
1
A
B
2