6 Mechanical conversion
6.4 Removing the old diaphragm unit
A
B
C
20
Converting Instructions "ORTHOPHOS XG 3D
6.4
Removing the old diaphragm unit
1. Pull cables L11 and L12 off connectors X501 (L11) and X101 (L12)
Removing the old diaphragm unit
Removing the diaphragm unit: XG5/XG 3D Ready
on board DX61.
2. Loosen the upper screw (A) and the two lower screws (B) and remove
the diaphragm unit.
IMPORTANT
DX61
If the two lower screws are not accessible, push the lower slider (C)
upward manually. If the slider is in such an unfavorable position that it
cannot be pushed upward:
– remove board DX61,
– loosen the four screws on the motor (D) and raise it with the spindle.
NOTE: The removed diaphragm unit is no longer required for reassembly.
D
Dispose of the old diaphragm unit in accordance with the regulations
applicable in your country [ → 48].
DX61
Sirona Dental Systems GmbH
ready
to ORTHOPHOS XG 3D" ORTHOPHOS XG
D3352.092.08.05.02 079.2012
63 14 541 D3352