Descargar Imprimir esta página

Sirona ORTHOPHOS XG 3Dready Manual De Instrucciones página 78

Juego de reequipamiento orthophos xg 3dready a xg 3d
Ocultar thumbs Ver también para ORTHOPHOS XG 3Dready:

Publicidad

Idiomas disponibles

Idiomas disponibles

6 Mechanical conversion
6.7 Attaching the ferrite cores Converting Instructions "ORTHOPHOS XG 3D
24
6.7
Attaching the ferrite cores
Attaching the ferrite cores
DANGER
Potentially lethal shock hazard!
Switch off the power supply connection before the power cable (D) is
disconnected or connected.
DANGER
Potentially lethal shock hazard!
Be sure to connect the second protective ground wire (C).
A
1. Fit the HA motor ferrite core (A) to the motor cable and attach this with
a cable tie to the sheet metal as shown in the figure to the left.
2. Loosen the three cores of the power cable (L = brown, N = blue, PE
= green/yellow) from the right side of the power line terminal. From
underneath, thread the cores twice through the ferrite core mains
filter (B) (in a loop).
3. Loosen the second, additional protective ground wire (C) (green/
yellow) from the stand.
4. Feed the second protective ground wire through the ferrite core (B)
from underneath.
5. Secure the cores with a cable tie to the ferrite core (B) as shown in
the drawing (right).
6. Connect the cores of the power cable (L, N and PE) to the power line
terminal once more.
7. Re-attach the second protective ground wire (C).
Sirona Dental Systems GmbH
ready
to ORTHOPHOS XG 3D" ORTHOPHOS XG 3D
D3352.092.08.05.02 079.2012
ready
D
B
C
63 14 541 D3352

Publicidad

loading