6 Mechanical conversion
6.9 Laying the light barrier and motor cables on the ring in the sliding plate brackets Converting Instructions "ORTHOPHOS XG 3D
Replacing old cable fixings with sliding
plate
Installing cables L20 and L21
26
6.9
Laying the light barrier and motor cables on the
ring in the sliding plate brackets
Laying the light barrier and motor cables on the ring in the sliding plate brackets
Condition before conversion
1. Remove all old (i.e. already present on the unit) cable fixings (A).
2. Insert the sliding plate (B) behind the edge of the rotary ring, push it
up, and clip it (C) to the rotary ring at both sides.
Condition following conversion
3. Use the two screws (D) to screw the sliding plate firmly onto the rotary
ring.
4. Remove the cables from the old cable fixings and lay them in the
brackets (E) of the sliding plate in the same order as before.
Only if not yet laid:
➢
Lay cables L20 (1x) and L21 (2x) (included in the upgrade kit) in the
sliding plate brackets.