7 Updating the unit software and configuring the unit
Saving data to board DX88
40
Converting Instructions "ORTHOPHOS XG 3D
1. Call service routine S009.8 (saving/restoring DX88 data) and execute
the routine (see ORTHOPHOS XG 3D Service Manual, section
"Service → S009: Test step 8").
2. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
The unit software has now been updated and the unit is configured.
Following a restart, only error message E1 11 20 (unit is not
calibrated) should appear.