6.5
Removing the old X-ray tube assembly
Removing the old X-ray tube assembly
1. Loosen the four screws (A) and remove cover plate (B) incl. the cable
shielding (L3). CAUTION! Also pull cable L3 off connector X3 and the
ground cable off connector X304 on board DX6.
Tip: The ferrite core and cable shielding can remain on the cover
plate.
2. Detach cables L5 and L6 from the rubber grommets and pull the
cables out of sockets J6 (L5) and J2-J3 (L6) on board DX6.
6 Mechanical conversion
ready
6.5 Removing the old X-ray