Descargar Imprimir esta página

Sirona ORTHOPHOS XG 3Dready Manual De Instrucciones página 77

Juego de reequipamiento orthophos xg 3dready a xg 3d
Ocultar thumbs Ver también para ORTHOPHOS XG 3Dready:

Publicidad

Idiomas disponibles

Idiomas disponibles

Sirona Dental Systems GmbH
Converting Instructions "ORTHOPHOS XG 3D
63 14 541 D3352
D3352.092.08.05.02 079.2012
ready
to ORTHOPHOS XG 3D" ORTHOPHOS XG 3D
6.6
Removing the Pan sensor holder
Removing the Pan sensor holder
Removing the Pan sensor holder: XG5/XG 3D Ready
C
D
E
1. CAUTION! Cable L13!
Loosen the three screws (C) and remove the sensor holder.
2. Loosen the two screws (D) and remove the "small sensor holder"
cover.
3. Loosen the three screws (E) and remove the connection socket.
4. Remove cable L13 from the shield terminal (F) at the rear of the
connection socket.
5. Loosen the connector screw connection on the front side of the
connection socket and thread the connector to the rear through the
connection socket and the sensor holder.
NOTE: The removed sensor holder is no longer required for reassembly.
Dispose of the sensor holder in accordance with the regulations
applicable in your country [ → 48].
6 Mechanical conversion
ready
6.6 Removing the Pan sensor
23