Sirona Dental Systems GmbH
Converting Instructions "ORTHOPHOS XG 3D
63 14 541 D3352
D3352.092.08.05.02 079.2012
ready
to ORTHOPHOS XG 3D" ORTHOPHOS XG 3D
8
Adjusting/calibrating the unit
Adjusting/calibrating the unit
1. In SIDEXIS XG, call the
"Utilities"
"Constancy test..."
/
"Service exposure "
calibration"
/ Password request (see also Service Manual, chapter
"Adjusting/calibrating the unit")
IMPORTANT
Before you begin adjusting or calibrating the unit, you must reset the 3D
adjustments.
2. In the structure tree, under
adjustment"
element.
"Reset values"
3. Click on the
IMPORTANT! After resetting the values, the parameter C1 must
have a value of "88".
4. Now perform a complete unit adjustment/calibration, as described in
the ORTHOPHOS XG 3D Service Manual, "Unit adjustment/
calibration" chapter.
8 Adjusting/calibrating the unit
ready
"Adjustment/calibration"
"3D"
"Select X-ray device"
/
/ (
"Select X-ray component"
/ (
"3D Reset adjustment"
button.
menu:
) /
"Adjustment/
) /
"Reset
, click on the