3.3 Torque Values of the Screw Connections
INFORMATION
The torque values for additional as well as optional system components (in item 3 below) are indicated in the Instruc-
tions for Use of those products.
15 Nm
Using the 710D4 Torque Wrench with the 710Y2=5 Hex Bit, turn the screws alternatively to the prescribed torque so
that the torque is gradually increased:
1 Tube adapter: 15 Nm
2 Tube clamp: 8 Nm
Fitting for short residual limb
3 Rotation adapter or sliding adapter: 15 Nm
Fitting for long residual limb
4 Lamination anchor with threaded connector: 10 Nm
3.4 Charging the C-Leg compact
INFORMATION
Advise your patients of the information in this section.
CAUTION
Incorrect behaviour while charging the joint. If the patient walks while the battery charger is connected to the
prosthesis, he can get caught on the cable and fall.
Take off the prosthesis prior to charging.
CAUTION
Incorrect handling while charging the joint. If the joint is not flexed completely during charging, the charging
plug can get damaged and the joint can no longer be charged. This can lead to insufficient power supply to the elec-
tronics of the joint and result in undefined conditions. This can cause the patient to fall.
During charging, keep the C-Leg compact fully flexed.
INFORMATION
Prior to charging with the 4E50-* Charger, read the corresponding 647G262 Instructions for Use.
34 | Ottobock
8 Nm
1
1
15 Nm
2
2
10 Nm
3
3
C-Leg compact Prosthetic System
4
4