GENERAL WARNINGS AND PRECAUTIONS
10
WARNINGS (CONTINUED)
• The Intelect® Transport 2 may be susceptible to Electro-Static Discharge (ESD) at greater than ±6 kV when first grasping the
Ultrasound applicator. In the event of such a discharge, the Intelect® Transport 2 may display a permanent error. The Intelect®
Transport 2 will terminate all active outputs (stim, ultrasound,), automatically place the unit in a safe state.Do not turn the unit on or
off while it is connected to the patient.
• To prevent Electro-Static Discharge (ESD) at greater than ±6 kV:
» Grasp and hold the Ultrasound prior to starting treatment. If the applicator must be put down prior to completion of treatment,
stop the current treatment first and then place the applicator in the holder.
» Maintain humidity in the use environment to at least 50% relative humidity.
» Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, DJO recommends implementing
additional controls to maintain relative humidity to at least 50%.
» Communicate these ESD-precautionary procedures to healthcare staff, contractors, visitors, and patients.
• In the event that an Error message or Warning appears beginning with a 2 or 3, immediately stop all use of the unit and contact
the dealer or DJO, LLC for service. Errors and Warnings in these categories indicate an internal problem with the unit that must be
tested by DJO, LLC or a Field Service Technician certified by DJO, LLC before any further operation or use of the system.
» Use of a unit that indicates an Error or Warning in these categories may pose a risk of injury to the patient, user, or extensive
internal damage to the system.
• Disconnect the system from the power source and battery before attempting any maintenance, installation, removal, or
replacement procedures to prevent electrical shock and possible damage to system.
• Device is not designed to be used in oxygen rich environment, Explosion hazard if the device is used in the presence of flammable
anesthetic mixture with air, oxygen, or nitrous oxide.
• Do not reverse the polarity of the battery pack. Doing so can increase the individual cell temperature and cause cell rupture or
leakage.
47979 INTELECT TRANSPORT 2 COMBO (4778) AND ULTRASOUND (4776) USER MANUAL