OPERATION
STARTING, STOPPING, AND INTERRUPTING ULTRASOUND THERAPY
To apply ultrasound therapy, do the following:
1.
Ensure the ultrasound applicator is plugged in.
2.
On the Home screen, use the Up and Down Arrow buttons
to highlight Ultrasound.
3.
Press the Enter button.
The Ultrasound parameters screen displays.
4.
Press the Up and Down Arrow buttons to highlight
Frequency.
5.
Press the Enter button to select 1 or 3 MHz.
When the button is pressed, the frequency will toggle from
1 to 3.3 MHz and back again as long as the button is being
pressed (excluding the 1 cm
An audible tone will be heard when changes are made.
2
NOTE: With 2 cm
and 5 cm
1 to 3.3 MHz results in no change in power. When using
2
a 10 cm
head with greater than 10 watts, changing from
1 to 3.3 MHz reduces power to 10 watts.
6.
Press the Up and Down Arrow buttons to highlight Duty
Cycle.
7.
Press the Enter button to display the Duty Cycle menu.
8.
Press the Up and Down Arrow buttons to highlight 10%,
20%, 50% or Continuous duty cycle.
When the buttons are pressed, an audible tone will be
heard as the duty cycle toggles through the options on
the LCD.
9.
Press the Enter button to accept the selection.
10. Press the Down Arrow button to highlight Display.
11.
Press the Enter button to select either Watts or W/cm
When the button is pressed, the display will toggle from
2
Watts to W/cm
and back again as long as the button is
being pressed.
47979 INTELECT TRANSPORT 2 COMBO (4778) AND ULTRASOUND (4776) USER MANUAL
2
sound head).
2
sound heads, switching from
2
.
12. Press the Up and Down Arrow buttons to select the manner
in which you want to be notified when the sound head
becomes uncoupled.
The following list details these options:
• Pause timer & beep - the timer stops and the unit
beeps once.
• Pause timer, no beep - the timer stops but the unit does
not give an audible tone.
• Run timer & beep - the timer continues to count down and
the unit beeps once.
• Run timer, no beep - the timer continues to count down,
but the unit does not give an audible tone.
NOTE: When the applicator becomes uncoupled during
treatment, it continues to distribute ultrasound energy.
13. Press the Enter button to accept the selection.
14.
Press the Time button and raise or lower treatment time
using the up and down arrows.
15. Press the INTENSITY button and raise or lower the unit's
output using the up and down arrows.
16. Press the Start button. The unit will beep 5 times and the
ultrasound power will distribute the selected output.
NOTE: When treatment time has expired a tone will sound
three times.
17.
The therapy can be interrupted at any time by pressing the
STOP or PAUSE buttons. When the STOP button is pressed,
the applicator stops emitting ultrasound energy, and the unit
returns to the default settings. To resume therapy, press the
Start button.
When the PAUSE button is pressed, the timer pauses and
the applicator stops emitting ultrasound energy. To resume
therapy, press the PAUSE button again.
18. The parameters of the therapy can be changed at any time
during the therapy session by pressing the appropriate
button.
19. After therapy is complete, wipe excess ultrasound gel from
the patient's skin and the sound head.