Descargar Imprimir esta página

Chattanooga Intelect Transport 2 Manual Del Usuario página 28

Ocultar thumbs Ver también para Intelect Transport 2:

Publicidad

Idiomas disponibles

Idiomas disponibles

28
SYSTEM
ULTRASOUND PATIENT PREPARATION
1.
Examine the skin for any wounds and clean the skin.
Applicator Preparation and Use
2.
Clean applicator before each therapy session with warm
soapy water, check the applicator has no cracks prior
to use.
3.
Liberally apply transmission gel to the treatment area on
the patient.
4.
Move the applicator during therapy session in a circular
motion. The area treated should be:
• Twice the diameter of the applicator
2
• For 5cm
US applicator: three times the diameter of the
applicator if output power > 4 W, Continuous mode.
The applicator should always be held by the grip and not by the
Ultrasound Applicator head.
5.
If US Coupling is "On", the Applicator is properly coupled
to the patient and administering ultrasound when the LED
is constantly illuminated. If the applicator head becomes
uncoupled the LED on the head will flash. If "US coupling"
setting is ON, several beeps will be also heard until
the head is coupled again. Treatment time stops during
uncoupling.
NOTE: Ultrasound output will continue to be emitted in all US coupling
modes even if the applicator is uncoupled. The output power is reduced
to a very low level to prevent ultrasound head warming.
47979 INTELECT TRANSPORT 2 COMBO (4778) AND ULTRASOUND (4776) USER MANUAL
STARTING, STOPPING, AND INTERRUPTING
ELECTROTHERAPY
The operator is able to navigate between treatment options,
view parameter options on the display and make selections
by pressing the buttons on the control panel. The display
will provide continuous information during the treatments
concerning power and elapsed time. Parameters are adjusted
using control panel buttons on the front of the unit. The output
can be stopped by pressing the "PAUSE" or "STOP" buttons
located on the control panel.
WARNING
Do not turn the unit on or off while it is connected to the pa ent.
Note: Electrotherapy related manual items are for Combo
device only
To apply electrotherapy, do the following:
1.
Follow all appropriate procedures listed in the section
entitled "Electrotherapy Patient Preparation" beginning on
page 27.
2.
Turn system power "ON" by pressing the Power On/Off
button. The message "Initializing System" displays while the
unit goes through self diagnostics until the home screen
appears.
3.
Connect the Lead Wires to the appropriate electrodes. To
see a list of recommended electrodes and their preparation,
see page 26.
NOTE: Do not use unnecessary force to connect the
electrodes to the lead wires.
4.
Place the self adhesive electrodes on the sites prescribed
by a qualified practitioner. Make sure you press them firmly
on the patient's skin to ensure good conductivity.
5.
Depending on the type of waveform you intend to use
and the number of patients you intend to treat, insert the
Lead Wire into Channel 1, Channel 2, or both Lead Wire
Connections on the Accessory Panel.
EN

Publicidad

loading