Descargar Imprimir esta página

Chattanooga Intelect Transport 2 Manual Del Usuario página 45

Ocultar thumbs Ver también para Intelect Transport 2:

Publicidad

Idiomas disponibles

Idiomas disponibles

ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES (CONTINUED)
The Intelect® Transport 2 is intended for use in the electromagnetic environment specified below.
The customer or the user of the Intelect® Transport 2 should assure that it is used in such an electromagnetic environment.
Immunity Test
IEC 60601 Test Level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
outside ISM bands
6 Vrms
150 kHz to 80 MHz
in ISM bands
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
9-28V/m
in wireless bands
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
a) The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz
to 27,283 MHz; and 40,66 MHz to 40,70 MHz.
Guidance and manufacturer's declaration – electromagnetic immunity
Compliance Level
3 Vrms
a
6 Vrms
a
3 V/m
9-28V/m
Electromagnetic Environment - Guidance
The device is intended for use in a typical professional
healthcare facility environment.
Portable RF communications equipment should be used
no closer than 30 cm (12 inches) to any part of the Intelect®
Transport 2, including cables specified by the
manufacturer. Otherwise, degradation of the performance
of this equipment could result.
APPENDIX
45

Publicidad

loading