4. OPERATION PROCEDURES
Operation Stop
(1) Press the Run/Stop key to stop rotation.
(2) Turn jack handle counterclockwise to release
the jack. Lift slide panel slowly so that
evaporation flask comes out of the bath.
(3) Stop the pressure reducing device. Gently
rotate Refill Valve so that the color marking on
the stopcock comes upside. Return unit to
atmospheric pressure.
(4) Stop cooling water circulator and water/oil
bath to end the operation.
(5) Remove flask clip from evaporation flask.
(6) Take evaporation flask off Vapor Tube.
When it is hard to remove evaporation flask by
hand, follow the steps below.
①Attach flask removal tool to joint nut. (See P.14)
②Hold evaporation flask with one hand, and turn
joint nut counterclockwise. Flask removal tool
pushes evaporation flask from the joint.
(7) press in the lock pin with the thumb to lock the
rotary part. Turn joint nut counterclockwise with
the other hand. Flask removal tool will now
push the neck of Vapor Tube. Pull out Vapor
Tube.
(8) Support receiving flask on the bottom, and
turn the dial inside ball joint clamp with the
other hand to release the lock. Remove the
clamp and receiving flask.
(9) Remove condenser fixing band.
* Only for vertical condenser (type F)
(10) Hold cooling condenser by hand and turn
condenser nut counterclockwise, then remove
cooling condenser.
Refill Valve
Color marking
Jack handle
Flask removal tool
Flask removal tool
Lock pin
Joint nut
Condenser nut
Turn counterclockwise to loosen
28
Run/Stop key
Flask clip
Joint nut
Flask removal tool
DE
EN
FR
ES
IT