Descargar Imprimir esta página

KNF Yamato RE212-G Manual De Instrucciones página 147

Rotavapor

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 258
3. PRE-OPERATION PROCEDURES
Installation Precautions
Use unit in fume hood whenever processing harmful solvents.
A ventilation hood must be installed for processing harmful solvents at constant
temperature.
Also obtain the safety data sheet (SDS) for safe use, and handle with extreme care.
【BO312-G】
Silicon oil when heated at more than 150°C will gradually generate trace amount of
formaldehyde. Place unit in fume hood or provide good ventilation for safe operation.
Heating medium
【BM312-G】
For water only * Using a fluid other than water may result in fire or equipment malfunction.
【BO312-G】
For water and oil * Maximum operating temperature is up to 90 °C for water, and 180°C
for oil. Use heat-resistant dimethyl silicon oil for open system heat transfer only, and
kinematic viscosity of 50mm
Recommended: KF-96-50cs silicon oil by Shinetsu Science Industries Co., Ltd.
Appearance
Kinematic viscosity
(25°C)
Specific gravity (25°C)
Volatile content
(150 °C/24 h)
Viscosity temperature
coefficient (V.T.C)
Pour point
Flash point
Specific Heat (25 °C)
Thermal conductivity
(25 °C)
Expansion coefficient
(25-150°C)
Deterioration rate of silicone oil varies depending on the operating temperature. For more
information on the silicone oil characteristics, contact silicone oil maker at the time of
purchase.
Use KF-96-50cs from Shinetsu Science Industries Co., Ltd. or its equivalent oils. KF-96
series silicone oil includes various types of viscosity. Note that low-viscosity oils have low
heat-resistance, and high-viscosity oils may cause local heating, possibly resulting in fire.
2
/s (cSt) or less.
Clear and colorless
2
50 mm
/s
0.960
0.5% or less
0.59
-50 °C or less
310 °C or more
1.5 J/g・°C
0.15 W/m・°C
0.00096 cc/°C
12

Publicidad

loading