KNF Yamato RE212-G Manual De Instrucciones página 152

Rotavapor

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 258
4. OPERATION PROCEDURES
Calibration Offset
Calibration offset is a function which can correct for any differences discovered between actual liquid
temperature and the temperature displayed on the control panel.
Offset function can correct to either the positive or negative side of the entire unit temperature range.
・Run unit at desired temperature. Once temperature has risen and stabilized, gauge liquid
temperature with a thermograph.
・Check the difference between the set temperature and the actual liquid temperature.
Setting range: -5.0 to +5.0 °C
Default setting is "0.0 °C"
 Setting change can also be made during operation.
Example
Actual temperature is 2 °C lower than control panel temperature reading of 60 °C.
Temperature reading can be calibrated by entering a calibration offset value of -2.0 to compensate
against the actual temperature deficiency of 2 °C.
If the initial temperature reading was 60°C, it will read 58°C after offset calibration, and be brought
into agreement with actual liquid temperature.
1. Enter user setting
2. Change offset value
3. Exit use setting
Turn power ON (|) and press
while temperature reading is on the screen.
Unit enters user setting.
Press 4 sec.
1. Select "CAL" using the ∧∨ keys.
Temperature display: "CAL" flashes.
2. Press
Temperature display: Current set value flashes.
3. Enter a value that brings set temperature and liquid
temperature into agreement, using the ∧∨ keys.
4. Press
Temperature display: Set value is shown for about one
After completing the setting, press and hold
for two seconds. Display reverts to initial screen and
shows temperature reading.
Press 2 sec.
key.
key to finalize.
second.
After completion, the screen returns to step 1.
17
key for four seconds
key
DE
EN
FR
ES
IT