4. OPERATION PROCEDURES
User Setting
List of user setting items
Press and hold
Press
key again to make setting on the selected item.
Holding down
key operation for about two minutes, will discard the changes, and the display returns to previous
screen.
Only calibration offset function "CAL" can be set or altered during operation (Run/Stop lamp ON).
The other items must be set during standby.
Panel item
Calibration offset
Auto-resume function
Overshoot alert
LED brightness setting
key for four seconds to show user setting. Select an item by using ∧∨ keys.
key for two seconds while the user setting item is displayed or leaving unit without
Calibration offset is a function which can correct for any
differences discovered between actual liquid temperature and
the temperature displayed on the control panel.
Offset function can correct to either the positive or negative side
of the entire unit temperature range.
Setting range: -5.0 to +5.0 °C
Default setting is "0.0 °C"
Select operation for the time power is restored.
OFF: Unit goes into idle at power recovery.
ON: Unit automatically reverts to status just before power
loss and begin operation once again from that point.
Default setting is "OFF"
When temperature reading goes over "set temperature + alert
setting value (°C)", the readout begins flashing to alert an
overshoot.
Alert setting range: 1-50 °C
Default setting is "50 °C"
Change the LED brightness of the control panel.
The brightness can be set in 8 levels from 0 to 7.
Setting range: 0-7
Default setting is "3"
Description
16
Page
P.17
P.18
P.19
P.20