High Intensity Light - Eye Damage
• Do not press activation button until properly positioned intraorally.
• Be sure everyone within the operating arena (patients, operators, assistants) is
wearing appropriate protective filtering eyewear.
• Do not look directly at the light while it is activated.
5. Adjust the LED tip: The LED tip is rotatable by 360°, thus the position of light emission
may be individually adjusted. The LED tip should be positioned as closely as possible to the
restoration. Avoid shadowing (e.g. by metal matrix or parts of the cavity) by angulating the
long axis accordingly. Fix correct position (e.g. with finger tip).
6. Polymerization
Briefly press the ON/OFF key
One Audible beep will be emitted. Handpiece will vibrate once.
7. SmartLite Pro is preset for 10 second cycles. To stop curing light before the end of the 10
second cycle, press ON/OFF key
a curing time longer than 10 seconds, repeat activation by pressing the ON/OFF key
after the end of each 10-second curing cycle
8. Curing times for a range of Dentsply Sirona materials are outlined in the Curing Guide
provided. When using other products, please refer to respective product's Instructions for
Use and apply curing times specified for 1000 mW/cm
distance to surface of material to be cured is larger than 4 mm.
9. To utilize a different application during the same treatment visit, the probe tip must be
changed. Remove light shield and barrier sleeve. Use moderate force to pull the probe tip
from the handpiece
the tip clicks into place while slightly rotating at the same moment
apply a new barrier if damage is suspected. Reattach the light shield.
10. Clean, disinfect and prepare contaminated handpiece and used tip(s) for reuse according
to section 4, Hygiene.
3.3 Operation – Illumination/Transillumination
1. Select the Transillumination (Illuminate) tip. Attach the tip to the handpiece by firmly
pressing tip into handpiece opening
2. Protect Illumination/Transillumination tip from gross debris by applying single use SmartLite
Pro barrier protection sleeve. After insertion turn the tip at least 180° to wrap sleeve around
tip for easier handling. Make sure that the lens is not blocked by crinkles or seam of the
sleeve.
Cross-contamination
• Ensure the disposable polyethylene FDA-cleared barrier protection sleeve has
been correctly applied over the entirety of the LED tip and handpiece before
beginning a procedure.
• The Polyethylene barrier Sleeve does not replace cleaning and disinfection of
the dental instruments. Please clean and reprocess dental instruments after each
patient as described in Section 4, Hygiene and Maintenance
High Intensity Light - Heat Damage
• Do not use curing tips for illumination or transillumination.
• Do not allow the tip to contact soft tissue for extended times
3. For visualizing anterior structures, briefly press the ON/OFF key
activate the lower output setting. To visualize posterior structures, briefly press the ON/OFF
key
1.1
a second time, which will activate the higher output setting.
4. For transillumination, apply the tip to the cervical area. Proximal defects are best visualized
by placing the tip towards the interproximal. Rotating the tip slowly will provide multiple
exposures of areas where caries is suspected. Cavitated areas typically appear as darkened
shadows within the tooth structure. Posterior teeth may also be illuminated by applying the
tip to the occlusal area so that vertical or horizontal cracks in enamel appear as dividing line
between differently illuminated sections.
5. When complete, press the ON/OFF key
12
1.1
to activate light.
1.1
at any time. To cure a restorative material that requires
C1
. Place the desired probe tip on the handpiece
C2
. Double provided curing time when
2
while slightly rotating at the same moment.
1.1
a third time, which will end the cycle.
C2
and press until
C3
. Reapply barrier, or
1.1
a single time, which will
1.1
C3