Packaging
No particular requirements.
Sterilization
Sterilization is not allowed. No methods have been validated.
Do not subject components to Steam autoclaving or liquid chemical sterilant immersion.
Component damage will occur.
Drying
Wipe the devices dry with a sterile, clean, lint-free cloth. Allow the components to fully
air dry before storage.
Maintenance,
Visually inspect to ensure that all contamination has been removed.
Inspection and
Visually inspect power supply and cord for damage.
Testing
Components that are damaged, worn, or distorted such as the O-rings should be
discarded and replaced.
See maintenance section below for additional recommended maintenance and testing.
Storage
Store the SmartLite Pro light handpiece, tips and charging base at room temperature,
away from moisture or excessive humidity.
Additional
Reassemble for Use as described above in step-by-step Instructions.
Information
Manufacturer
For areas outside the United States, contact your local Dentsply Sirona representative.
Contact
Instructions for Cleaning, Disinfecting and Sterilizing SmartLite Pro Light Shield
Warnings
• These instructions are for use ONLY for the light eye protection Shield. The
handpiece, tips and charging base should be disinfected according to the
procedures in the "Instructions for Cleaning and Disinfecting SmartLite Pro
Light Handpiece, Tips, Charging Base" section above.
• The SmartLite Pro Shield should be removed and cleaned/disinfected/
sterilized as outlined below. The SmartLite Pro Light handpiece, tips and
charging base are not sterilizable by autoclave.
• High level disinfection is appropriate for the light shield.
• Steam autoclaving sterilization is recommended for the light shield.
• Do not allow the device to exceed 134°C.
Limitations on
• Repeated reprocessing has minimum effect on these instruments. End of life
reprocessing
is normally determined by wear and damage due to use.
• Device can be reprocessed for at least 60 times.
• Cold liquid immersion disinfection/sterilization, chemical vapor sterilization,
and dry heat sterilization methods have not been tested or validated for
efficacy and are not recommended for use.
Initial treatment
• Use moderate force to pull the SmartLite Pro Light Shield from the
at the point
handpiece.
of use
• Remove protective barrier and discard according to local regulations.
• Use a new, clean pair of examination gloves.
• Remove excess soil with disposable cloth / paper wipe in combination with a
pH-neutral, phosphate-free cleaning solution
(e.g.: Dr. Schumacher Instru Plus [3%])
• Reprocess SmartLite Pro Light Shield as outlined below. Reprocess
handpiece, tips and charging base as outlined in the "Instructions for
Cleaning and Disinfecting SmartLite Pro Light Handpiece, Charging Base"
section above.
• It is recommended that the device be reprocessed as soon as is reasonably
practical following use.
• Start reprocessing within 1 hour after use.
Preparation
Always disassemble eye protection shield from handpiece before processing.
before cleaning
15