6.6 Disposal of Unit
This device is provided with a lithium-ion phosphate battery. Device and battery must not
be disposed of in normal domestic waste. For environmental reasons, dispose of device and
battery according to local environmental guidelines or regulations.
6.7 Electromagnetic Compatibility Precaution
This information is required by the 4th edition of IEC 60601-1-2.
• The SmartLite® Pro Light needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided in this manual.
• Portable and mobile RF communications equipment can affect the SmartLite® Pro Light.
• The use of accessories, transducers and cables other than those specified by Dentsply
Sirona, may result in increased emissions or decreased immunity of the SmartLite® Pro Light.
• The SmartLite® Pro Light should not be used adjacent to or stacked with other equipment
and that if adjacent or stacked use is necessary, the SmartLite® Pro Light should be
observed to verify normal operation in the configuration in which it will be used.
• Per IEC 60601-1-2, no additional environmental operating conditions are required for normal use
Guidance and manufacturer's declaration – electromagnetic emissions
The SmartLite® Pro Curing Light is intended for use in the electromagnetic environment specified below. The customer
or the user of the SmartLite® Pro Light should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR11
RF emissions
CISPR11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The SmartLite® Pro Curing Light is intended for use in the electromagnetic environment specified below. The customer or
the user of the SmartLite® Pro Curing Light should assure that it is used in such an environment.
IMMUNITY Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
20
Compliance
Electromagnetic environment - guidance
The SmartLite® Pro Curing Light uses RF energy only for its internal
Group 1
function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
Class B
The SmartLite® Pro Curing Light is suitable for use in all
Class A -
establishments, including domestic establishments and those
Complies
directly connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Complies
IEC 60601
Test level
± 8 kV contact
± 2, ± 4, ± 8, ± 15 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 0.5, + 1 kV line(s) to
line(s)
± 0.5, ± 1, ± 2 kV Line(s)
to earth
Compliance level
± 8 kV contact
± 2, ± 4, ± 8, ± 15 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 0.5, + 1 kV line(s) to
line(s)
± 0.5, ± 1, ± 2 kV Line(s)
to earth
Electromagnetic environment -
guidance
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.