Multicheck6010
4.8 Earth Resistance
4.8.1 Earth Resistance (Re) - 3-wire, 4wire
How to perform Earth Resistance measurement
Step 1
Select the Earth Resistance function with the function selector FCT key and select the
Re mode with the ▲▼ and ◀▶ navigation keys. The following menu will be displayed:
Step 2
Set the following limit value:
• Limit: : limit resistance value using the ▲▼ and ◀▶ navigation keys.
Step 3
Follow the connection diagram shown in figures 4.36 to perform the Earth Resistance
measurement with 4 wires and the connection diagram shown in figures 4.37 to
perform the Earth Resistance measurement with 3 wires (ES connected to E).
Figure 4 36: 4 wire connection diagram
Step 4
Check for any warnings and the online voltage/terminal monitor on the display before
starting the measurement. If everything is ok and the
After performing the measurement, the results appear on the display together with the
indication (if applicable).
Figure 4 35: Earth Resistance (Re) measurement menu
Figure 4 37: 3 wire connection diagram
is shown, press the TEST key.
- 82 -
EN
or