Continuity Resistance; Low R; Low Current Continuity; Rcd Testing - KPS Multicheck6010 Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Multicheck6010

6.2 Continuity resistance

6.2.1 Low R

Measuring range according to EN61557-4 is 0.16 Ω ÷ 1999 Ω.
Range (Ω)
0.1 ÷ 20.0
20.0 ÷ 1999
Open-circuit voltage
Measuring current
Test lead compensation
Number of possible tests
with a new set of batteries
Automatic polarity reversal of the test voltage.

6.2.2 Low current continuity

Range (Ω)
0.1 ÷ 1999
Open-circuit voltage
Corriente de cortocircuito
Test lead compensation

6.3 RCD testing

6.3.1 General data

Nominal residual current
Nominal residual current accuracy -0 / +0.1·IΔ; IΔ = IΔN, 2xIΔN, 5xIΔN
Test current shape
RCD type
Test current starting polarity
Voltage range
Resolution (Ω)
(0.10 Ω ... 19.99 Ω) 0.01 Ω
(20.0 Ω ... 99.9 Ω) 0.1 Ω
(100 Ω ... 1999 Ω) 1 Ω
5 VCC
min. 200 mA into load resistance of Ω
up to 5 Ω
up to 1400 (with 2300mAh battery cells)
Resolution (Ω)
(0.1 Ω ... 99.9 Ω)
0.1 Ω
(100.0 Ω ... 1999 Ω) 1 Ω
5 VCC
max. 7 mA
up to 5 Ω
10 mA, 30 mA, 100 mA, 300 mA, 500 mA, 650mA, 1000 mA
-0.1·IΔ / +0; IΔ = ½xIΔN
Sinusoidal (AC), CC (B), impulso (A)
general (G), selective (S, time-delayed)
0º ó 180º
93V-134V; 185V-266V; 45Hz-65Hz
Accuracy
±(3 % of reading + 3 digits)
±(5 % of reading)
Accuracy
±(5 % of reading + 3 digits)
- 87 -