Multicheck6010

6. Technical specifications

6.1 Insulation resistance

Insulation resistance (nominal voltages 50VDC,100 VDC and 250 VDC)
Range (MΩ)
0.1 ÷ 199.9
Insulation resistance (nominal voltages 500 VDC and 1000 VDC)
Range (MΩ)
0.1 ÷ 199.9
200 ÷ 999
Voltage
Range (V)
0 ÷ 1200
Nominal voltages
Open circuit voltage
Measuring current
Short circuit current
The number of possible tests
wuth a new set of batteries
Auto discharge after test
In case the instrument gets moistened the results could be impaired. In such case it is
recommended to dry the instrument and accessories for at least 24 hours.
Resolution (MΩ)
(0.100 ... 1.999)
0.001
(2.00 ... 99.99)
0.01
(100.0 ... 199.9)
0.1
Resolution (MΩ)
(0.100 ... 1.999)
0.001
(2.00 ... 99.99)
0.01
(100.0 ... 199.9)
0.1
(200 ... 999)
1
Resolution (V)
1
50VCC, 100 VCC, 250 VCC, 500 VCC, 1000 VCC
-0 % / +20 % of nominal voltage
min. 1 mA at RN=UNx1 kΩ/V
max. 15 mA
up to 1000 (with 2300mAh battery cells)
Accuracy
±(5 % of reading + 3 digits)
Accuracy
±(2 % of reading + 3 digits)
±(10 % of reading)
Accuracy
±(3 % of reading + 3 digits)
- 86 -