Visible damages
Damage to the power cord
Damage to the fuse holder
Loose parts
No operation
After longterm storage in an inappropriate environment ,
e.g. open air or high humidity.
Excessive transport stress
Operating conditions
The instruments are destined for use in dry clean rooms. Ope-
ration in an environment with high dust content, high humidity,
danger of explosion or chemical vapors is prohibited.
Operating temperature is 0 .. +40 degrees C. Storage or trans-
port limits are –10 .. +70 degrees C. In case of condensation
two hours are to be allowed for drying prior to operation.
For safety reasons operation is only allowed from 3 terminal
connectors with a safety ground connection or via isolation
transformers of class 2. The instrument may be used in any
position, however, sufficient ventilation must be assured as
convection cooling is used. For continuous operation prefer a
horizontal or slightly upward position using the feet.
Do not cover either the holes of the case nor the
cooling fins.
Nominal specs are valid after a warm-up period of min. 20
min. in the interval of +15 to +30 degrees C. Values without a
tolerance are typical of an average production instrument.
STOP
Warranty and Repair
HAMEG instruments are subject to a strict quality control. All
instruments are burned in for 10 hrs prior to shipment. By
intermittent operation almost all early failures are detected.
After burn-in a thorough test of all functions and of quality is
run, all specifications and operating modes are checked.
In case of reclamations during the two years warranty period
please contact the dealer from whom you purchased your
HAMEG instrument. Customers from the Federal Republic of
Germany may directly contact HAMEG for warranty processing
in order to speed up the procedure.
The proceeding of repairs during the warranty period is subject
to our terms of warranty which are available on our website
http://www.hameg.com
Even after expiry of the warranty period please do not hesitate
to contact our HAMEG customer service for repairs and spare
parts.
Return Material Authorization (RMA):
Before sending back your instrument to HAMEG do apply for
a RMA number either by fax or on the Internet: http://
www.hameg.de.
If you do not have suitable packaging for the instrument on
hand please contact the HAMAG sales department (Tel.: +49
(0) 6182/800 300, E-mail: vertrieb@hameg.de) to order an
empty original cardboard box.
Maintenance
The instrument does not require any maintenance. Dirt may
be removed by a soft moist cloth, if necessary adding a mild
detergent. (Water and 1 %.) Grease may be removed with
I m p o r t a n t
benzine (petrol ether). Displays and windows may only be
cleaned with a moist cloth.
Do not use alcohol, solvents or paste. Under no
circumstances any fluid should be allowed to get into
the instrument. If other cleaning fluids are used
STOP
STOP
damage to the lacquered or plastic surfaces is
possible.
Line voltage selector
The instrument is
destined for opera-
!
tion on 115 or 230 V
CAT II
mains, 50/60 Hz.
The proper line
Voltage
voltage is selected
with the
line
voltage selector. It
Selector
is necessary to
change the fuse ob-serving the proper values printed on the
back panel.

Change of fuse

The mains fuse
is accessible on the back panel. A change
of the fuse is only allowed after the instrument was
disconnected from the line and the power cord removed. Fuse
holder and power cord must not show any sign of damage.
Use a screw driver to loosen the fuse holder screw
counterclockwise while pressing the top of the fuse holder
down. The top holding the fuse will then come off. Exchange
the defective fuse against a correct new one. Any „repair" of a
defective fuse or brid-ging is dangerous and hence prohibited.
Any damages to the instrument incurred by such manipula-
tions are not covered by the warranty.
Type of fuse:
5 x 20 mm; 250V~, C;
IEC 127/III; DIN 41662
(DIN 41571/3).
Value
115 V: 200 mA slow blow
230 V: 100 mA slow blow
h i n t s
INTERNAL INSTRUMENT SUPPLY
115 230VAC / 50 60 Hz