2.1.2 Operating Warnings
• Unpack and inspect this device for damage prior to installation, setup, or use . Do not use this
device if it is damaged . Do not use this device if it malfunctions .
• This device is for use by personnel trained in endoscopic procedures .
• Do not use this device in the presence of explosive gases .
• Factory settings should only be used as guidelines . The surgeon is responsible for settings
pertaining to the surgical procedure .
• Failure to follow all applicable instructions may result in serious surgical consequences to the
patient .
• Any modifications to the system must be authorized by DePuy Synthes . Modifying the device
in a manner other than specified by DePuy Synthes may result in injury .Contact Customer
Service at +1 (800) 382-4682 for service requests .Outside the United States, contact your local
affiliate .
• The Camera Control System is equipped with a terminal for the connection of a Potential
Equalization Connector as required by Electrical Safety Standard IEC 60601-1 . The Potential
Equalization Connector is provided to prevent electric shock hazards by maintaining a constant
potential (voltage) to all exposed metal parts so a person cannot simultaneously make contact
with two different potentials . Use of the potential equalization connector should be determined
by the facility where it is installed based on the design of the electrical distribution system .
• If using the Camera Control System without the ITD cart, connect one end of the Potential
Equalization Connector to the terminal on the back of the Camera Control System (Figure 2,
Item 11) . Attach the other end to the bus bar of the electrical installation .
• If using the Camera Control System with the ITD cart, connect one end of the Potential
Equalization Connector to the terminal on the back of the Camera Control System (Figure 2,
Item 11) . Attach the other end to the ITD cart (Refer to IFU-115831) . Connect the cart's
equalization connector to the bus bar of the electrical installation .
• Accessories labeled "REUSABLE" must be processed only according to the recommended
procedures . Refer to the applicable Instructions for Use .
• The use of accessories, transducers and cables other than those specified, with the exception
of transducers and cables sold by the Manufacturer of the ME Equipment or ME System, may
result in increased emissions or decreased immunity of the ME Equipment or ME System .
• No part of the system can be serviced or maintained while the unit is in use .
• Do not look directly into the light source . The intense light may cause eye injury .
2.2
Precautions
• Read the instructions, cautions, and warnings provided with all System accessories before use .
Your sales representative can advise which accessories are used with the System .
• Test this equipment prior to installation or setup . Perform pre-use checks prior to each use .