4. Clean
• Prepare cleaning solution using neutral pH enzymatic detergent per
manufacturer's directions for correct temperature, water quality (i.e., pH,
hardness) and product concentration/dilution.
• With distal end pointing down spray and wipe using cleaning solution for a
minimum of 2 minutes.
• With distal end pointing down rinse device thoroughly using running cold
deionized, distilled or purified water for a minimum of 2 minutes while
manipulating knurled knob.
• Use a sponge, soft lint-free cloth, or soft bristle brush to remove all visible soil.
• Place the locking mechanism in the "SAFE" position prior to performing the
following step. See Figure 3.
Figure 3
• Hold handpiece with distal end pointing down and insert ACB wetted with
cleaning solution into distal end of handpiece. See Figure 4.
Warning: ACB is single use and should only be used to clean one handpiece
and associated attachments.
Caution: Do not immerse. Do not allow running water to enter distal end
of handpiece. Do not insert ACB beyond last bristle. Do not force ACB into
handpiece. Do not use pipe cleaners.
Figure 4
5. Rinse Thoroughly
• With distal end pointing down rinse device thoroughly using running hot
deionized, distilled or purified water for a minimum of 2 minutes.
• Place the locking mechanism in "SAFE" position prior to performing the
following step. See Figure 5.
• Hold handpiece with distal end pointing down and flush inside of distal end
using a syringe or pipette filled with hot deionized, distilled or purified water.
• Place knurled knob in "RUN" position. See Figure 5.
Figure 5
Caution: Do not use high pressure water or air. Do not immerse. Do not allow
running water to enter distal end of handpiece.
6. Inspect
• Visually inspect handpiece. Repeat Steps 2 thru 6 until no visible soil remains.
Warning: Manual pre-cleaning must be followed by mechanical/automated
cleaning.
6