1. Preparation
• Remove attachment and dissection tool from handpiece.
• Assemble all necessary supplies including Anspach Cleaning Brush
(ACB), Small Diameter Cleaning Brush (SDCB), sponge, lint-free cloth
and soft bristle brush.
Note: Red storage cap provided with Perforator Driver (CSR60) is to
prevent damage to shaft. Cap is removed for cleaning, washing and
sterilization. Cap should be used for non-sterile storage purposes only.
2. Rinse
• Rinse attachment under running cold deionized, distilled or purified
water for a minimum of 2 minutes.
• Use a sponge, soft lint-free cloth, or soft bristle brush to assist in
removing gross soil.
Caution: Do not use high pressure water or air. Do not rinse attachment
with saline solution.
3. Manipulate Moving Parts
• Manipulate all moving parts such as release sleeves under running
cold deionized, distilled or purified water to loosen or remove gross
soil.
4. Clean
• Prepare cleaning solution using neutral pH enzymatic detergent
per manufacturer's directions for correct temperature, water quality
(i.e. pH, hardness) and product concentration/dilution.
• Fully immerse attachment in cleaning solution in a suitable container
and manually agitate for 2 minutes.
• Use a sponge, soft lint-free cloth, or soft bristle brush to remove all
visible soil.
• Gently use ACB or SDCB wetted in cleaning solution to clean
cannulations of attachment. Brushes may be inserted from distal or
proximal end of attachment.
Warning: ACB and SDCB are single use and should only be used to
clean one set of attachments.
Caution: Do not force ACB or SDCB into or through attachment. Do not
use ACB or SDCB with MA-D20 and MA-DRIVER. For Minimal Access
Attachment Bearing Sleeves use SDCB only. Do not use pipe cleaners.
Note: "DO NOT IMMERSE" as marked on the attachment is for
Operating Room personnel only. SDCB may be inserted from distal or
proximal end of bearing sleeve.
5. Rinse Thoroughly
• Rinse device thoroughly using running hot deionized, distilled or
purified water for a minimum of 2 minutes.
• Use a syringe or pipette filled with hot deionized, distilled or purified
water to flush lumens or channels.
Caution: Do not use high pressure water or air. Do not rinse attachment
with saline solution.
6. Inspect
• Visually inspect attachment. Repeat Steps 2 thru 6 until no visible
soil remains.
Warning: Manual pre-cleaning must be followed by mechanical/
automated cleaning.