5.8 Mechanical cleaning/disinfecting
Note
The disinfector must be of tested and approved effectiveness (e.g. DGHM or
FDA approval or CE mark according to DIN EN ISO 15883).
Note
For thermal disinfection, always use fully desalinated (demineralized)
water and ensure that Ao is >3 000 for the process.
Note
The disinfector used for processing must be serviced and checked at regular
intervals.
Insert the product in its proper position in the Eccos® holder.
5.8.1 Tools
Mechanical alkaline cleaning and thermal disinfecting
Machine type: Single-chamber washer/disinfector without ultrasound
Stage
Step
I
Prerinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfecting
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized)
T
t
Water quality
[°C/°F]
[min]
< 25/77
3
55/131
10
>10/50
1
90/194
5
-
-
Chemical/Note
D–W
-
FD–W
BBRAUN HELIMATIC CLEANER alcaline
with tensides; application solution 0.5 %
FD–W
-
FD–W
-
-
according to disinfecting program
13