4.2.2 Uncoupling a tool

Hold tool 12.
Pull back slide 7 on craniotome handpiece 1.
Remove tool 12 from the craniotome handpiece.
4.3 Mounting the dura protector or depth stop and

holding sleeve on the craniotome handpiece

4.3.1 Mounting the dura protector (fixed) or the depth stop and

holding sleeve on the craniotome handpiece
Turn swivel ring 10/18/21 of dura protector (fixed) 9 or depth stop 15
or holding sleeve 20 in the direction of the arrow in symbol
the limit stop.
Insert dura protector (fixed) 9 or depth stop 15 or holding sleeve 20 in
craniotome handpiece 1 in such a way that the pin of the dura
protector (fixed) or attachment is seated in one of the recesses of the
craniotome handpiece.
Turn swivel ring 10/18/21 in the direction of the arrow in symbol
to the limit stop.
Make certain that swivel ring 10/18/21 is fully fastened to prevent it
from loosening due to vibrations during operation.
The dura protector (fixed)/attachment is locked now.
4.3.2 Mounting the dura protector (pivoted) on the craniotome
handpiece
Turn swivel ring 23 of dura protector (pivoted) 24 in the direction of
the arrow in symbol
as far as it will go.
Install dura protector (pivoted) 24 on craniotome handpiece 1.
Hold dura protector mount 22 and turn swivel ring 23 in the direction
of the arrow in symbol
Make certain that swivel ring 23 is fully fastened to prevent it from
loosening during operation.
Ensure that dura protector (pivoted) 24 can be turned freely and easily
on the craniotome handpiece.
The dura protector (pivoted) is locked now.
Note
The dura protector and craniotome reamer are mounted correctly if a
Y
narrow gap
is left between the dura protector foot and the craniotome
reamer.
6
as far as it will go.

4.4 Uncoupling of the dura protector/attachment

4.4.1 Uncoupling the dura protector (fixed)/attachment

Turn swivel ring 10/18/21 of dura protector (fixed) 9 or depth stop 15
or holding sleeve 20 in the direction of the arrow in symbol
the limit stop.
Remove dura protector (fixed) 9 or the attachment.

4.4.2 Uncoupling the dura protector (pivoted)

Hold dura protector mount 22 and turn swivel ring 23 of dura
protector (pivoted) 24 in the direction of the arrow in symbol
far as it will go.
to
Remove dura protector (pivoted) 24.
4.5 Coupling the craniotome handpiece to the HiLAN®
XS motor
Turn safety stop 5 on motor 4 to the "0"-position.
Turn threaded ring 2 of craniotome handpiece 1 in the direction of the
arrow in symbol
as far as it will go.
Install the craniotome handpiece on the motor in such a way that the
pin on the craniotome handpiece is seated in one of the recesses in the
hand piece coupling of the motor.
Turn threaded ring 2 in the direction of the arrow in symbol
far as it will go.
Make certain the threaded ring is fastened to its limit stop, because
otherwise it may come loose due to vibrations during operation.
The craniotome handpiece is locked now.
4.6 Uncoupling the craniotome handpiece from the
HiLAN® XS motor
Turn safety stop 5 on motor 4 to the "0"-position.
Turn threaded ring 2 in the direction of the arrow in symbol
far as it will go.
Detach craniotome handpiece 1 from the motor.
to
as
as
as